| Data Box |
Specifications |
Inputs: |
Pressure: |
2 differential pressure channels with auto-zeroing (APT-2). 3 to 8 differential pressure channels with auto-zeroing (APT-8). Switchable resolution of 0.1 Pa and 0.5 (Pa), with corresponding ranges of approximately +/-400 Pa and +/- 1,000 Pa, respectively. |
Voltage: |
8 differential analog DC voltage input channels. Switchable resolution of 0.1 mV and 1 mV, with corresponding ranges of approximately +/-400 mV and +/-4 V, respectively. Voltage channels are accessed via a bank of 8 six-contact phone jacks. Continuous 5V (regulated,+/-1%, max 50 mA) power available for optional sensors. |
Outputs: |
Automated fan speed control connection to Minneapolis Blower Door and Duct Blaster systems.
2 switchable digital outputs (5V regulated). |
Communication: |
RS-232 connector (9-pin) to interface to your PC at 9,600 baud. |
Power: |
Internal NiCd battery pack with AC power adapter. |
Accuracy: |
Pressure: |
+/- 1% of reading, or 2 x resolution, whichever is greater from 0-800 Pa (at 10 °C to 30 °C). +/- 2% of reading from 800-1,000 Pa. |
Voltage: |
+/- 0.1% of reading, or 2 x resolution, whichever is greater from 0-4 Volts (at 10 °C to 30 °C). |
Dimensions: |
4"x 8" x 1.5" (APT-2). 4" x 10"x 3" (APT-8). |
Temperature Sensor |
Sensing Method: |
Precision NTC Thermistor, 10k Ohms @ 25 °C. |
Accuracy: |
+/- 0.25 °C (0 to 75 °C),
+/- 0.5 °C (<0 °C, >75 °C). |
Range: |
-40 °C to 100 °C (-40 °F to 212 °F). |
Resolution: |
0.025 °C (0.045 °F) @ -40 °C (-40 °F).
0.05 °C (0.09 °F) @ 21 °C (70 °F).
1.0 °C (1.8 °F) @ 100 °C (212 °F). |
Time Constant: |
11 seconds in slowly moving air, 70 seconds in still air. |
Operating Temperature: |
-40 °C to 100 °C (-40 °F to 212 °F). |
Storage Temperature: |
-40 °C to 100 °C (-40 °F to 212 °F). |
Current Draw: |
11 µA @ -40 °C (-40 °F).
38 µA @ 0 °C (32 °F).
50 µA @ 100 °C (212 °F). |
Power Supply: |
Provided by APT data acquisition box. |
Long-Term Stability: |
3000 hrs @ 20 °C (68 °F):
+0.018 °C (+ 0.03 °F) typical.
2000 hrs @ 70 °C (160 °F):
+0.07 °C (+ 0.126 °F) typical. |
Housing: |
1 1/2" nominal length, 3/16" O.D. |
Lead Length: |
10' 4 conductor flat phone cable. |
Calibration Interval: |
1 Yr. recommended (calibration adjustments are made using TECLOG channel Parameters). |
Warranty |
1 year. |
Relative Humidity Sensor |
Sensing Method: |
Thermostet Polymer based capacitance relative humidity sensor. |
Accuracy: |
+/- 5% RH (0 to 60% RH),
+/- 8% RH at 90% RH (@ 25 °C, non-condensing).
|
Range: |
0 to 100% RH. |
Linearity: |
+/- 0.5% RH typical. |
Repeatability: |
+/- 0.5% RH. |
T90 Response Time: |
30 seconds in slowly moving air typical. |
Long-Term Stability: |
+/- 1% RH typical at 50% RH in 5 years. |
Hysteresis: |
+/- 0.8% RH maximum (included in accuracy). |
Temperature: |
Actual RH = Sensor RH / (1.093-.0012T) Compensation: where T=°F. |
- Effect on Zero Shift: |
+/- 0.007% RH/°C (Negligible). |
- Effect on Span: |
-0.22% RH/°C (<1% RH effect typical in occupied systems above 15 °C). |
Operating Temperature: |
-40 to 85 °C (-40 to 185 °F). |
Storage Temperature: |
-51 to 110 °C (-60 to 230 °F). |
Output: |
0.8 - 3.9 Vdcout (0 to 100%RH) for 5 Vdcsupply. |
Current Draw: |
200 µA typical @ 5 Vdc. |
Power Supply: |
Provided by APT data acquisition box. |
Housing: |
2" nominal length, 5/16" O.D. |
Lead Lengths: |
10' 4 conductor flat phone cable. |
Handling: |
Do not immerse in water or subject to prolonged condensation. |
Calibration Interval: |
1 Yr. recommended (calibration adjustments are made using TECLOG channel Parameters). |
Warranty |
1 year. |
Carbon Monoxide Sensor |
Sensing Method: |
Electrochemical (fuel cell). |
Accuracy: |
+/- 5% of reading (@ 20 °C, 50% RH, 1013 mBar). |
Range/Resolution: |
0 - 2,000 ppm / 0.5 ppm. |
Linearity: |
Linear (full range). |
Repeatability: |
+/- 1% of signal. |
T90 Response Time: |
< 30 seconds. |
Long-Term Stability: |
< 2% of signal/month. |
Temperature Effect on Span: |
0.5%/°C typical, 0.75%/°C maximum. |
Barometric Pressure Effect: |
0.02% signal/mBar. |
Maximum Zero Shift: |
3 ppm (20 °C to 40 °C). |
Typical Baseline Range: |
0 +/- 0.5 ppm. |
Operating Conditions: |
|
- Temperature |
-20 °C to 45 °C. |
- RH |
15 to 90%, non-condensing. |
- Pressure |
Atmospheric +/- 10%. |
Storage Temperature: |
0 to 20 °C. |
Output: |
1mV/ppm. |
Current Draw: |
0.5 µA typical @ 5 Vdc. |
Power Supply: |
Provided by APT data acquisition box. |
Expected Operating Life: |
2 years in air, minimum. |
Housing: |
3" x 2 1/2" x 1 1/2" anodized aluminum w/ RJ11 phone jack. |
Lead Lengths: |
10' 4 conductor flat phone cable. |
Handling: |
Avoid condensing conditions and exposure to >2,000 ppm. |
Calibration Interval: |
6 months recommended (if CO sensor is being used for safety purposes, more frequent calibration checks may be desired). |
Warranty |